Estudo comparativo das características microscópicas e macroscópicas entre limas originais e limas de contrafação

dc.contributor.advisorAzul, Ana Mano
dc.contributor.advisorPereira, Mário Rito Dias
dc.contributor.authorFurtado, Sofia Magalhães de Moura e Mota
dc.date.accessioned2022-01-14T13:29:17Z
dc.date.available2024-12-31T01:30:28Z
dc.date.issued2021-12
dc.descriptionDissertação para obtenção do grau de Mestre no Instituto Universitário Egas Monizpt_PT
dc.description.abstractObjetivos: Este estudo teve como objetivo avaliar as propriedades microscópicas e macroscópicas de 14 marcas comerciais diferentes de limas endodônticas do tipo K-File para verificar se seguem as normas ISO e ANSI/ADA, bem como outras propriedades não contempladas nas normas. Materiais e Métodos: Conjuntos de limas de 1ª série do tipo K-File de aço inoxidável de 13 marcas originais (VDW, Dentsply Ready Steel, Falcon, Torino, Henry Schein, Mani, M-Precise, SybronEndo, Thomas, Komet, Smartdent, UDG e R&S) e de uma marca de contrafação (Dentsply Maillefer) (n = 84), foram selecionadas para avaliação das propriedades macroscópicas, com recurso a um microscópio cirúrgico Zeiss OPMI Pico, e propriedades microscópicas, com recurso ao MEV. Posteriormenteu, um total de 616 fotografias adquiridas através do MEV e do Microscópio Cirúrgico, foram trabalhadas no Adobe Photoshop CC 2019, para se substituir o fundo de cada imagem por um fundo preto de forma a facilitar a visualização das mesmas. Para se quantificar os valores das propriedades avaliadas, foi utilizado o programa ImageJ sempre com uma escala calibrada que está presente em cada uma das imagens. Para cada parâmetro analisado foram efetuadas 3 medições diferentes sendo que o valor apresentado corresponde à média aritmética dos valores medidos. Resultados: Verificaram-se diferenças significativas entre todas as limas analisadas quando comparadas entre si, com as normas ANSI/ADA e ISO e com outras variáveis. Conclusão: Na pesquisa bibliográfica efetuada não foram encontrados estudos que comprovem a semelhança entre limas de contrafação e limas originais, sendo como tal relevante avaliar as características de cada uma e verificar se correspondem aos ditos padrões. Com este estudo concluímos que nenhuma das limas analisadas cumpre as normas ISO e ANSI/ADA, e por isso o controlo de qualidade deve ser avaliado de forma a conseguir-se obter uma padronização dos instrumentos endodônticos que estão disponíveis no mercado.pt_PT
dc.description.abstractPurpose: The purpose of this study was to evaluate the microscopic and macroscopic properties of 14 different brands of endodontic K-Files in order to verify if they followed ISO and ANSI / ADA specifications, and assess other properties not mentioned on these specifications. Materials and Methods: Stainless steel K- Files (sizes 15-40) from 13 original brands (VDW, Dentsply Ready Steel, Falcon, Torino, Henry Schein, Mani, M-Precise, SybronEndo, Thomas, Komet, Smartdent, UDG and R&S) and 1 counterfeit brand (Dentsply Maillefer) (n = 84) were selected to evaluate their macroscopic and microscopic properties, by using a Zeiss OPMI Pico surgical microscope and SEM, respectively. Subsequently, a total of 616 photographs were acquired either through SEM or Surgical Microscope, and afterwards worked on Adobe Photoshop CC 2019, in order to replace the background of each image with a black one for better visualization. In order to quantify the values of the studied properties, the ImageJ program was always used with a calibrated scale that is present in each of the images. For each analyzed parameter, 3 different measurements were taken, being the shown value, the arithmetic mean of the measured values. Results: There were significant differences between all analyzed files, when compared to each other, to ANSI/ADA and ISO standards, and other variables. Conclusion: In the bibliographical research carried out, no studies were found to prove the resemblance between counterfeit files and original files, making it relevant to evaluate the characteristics of each file and check if they are within said standards. With this study, we concluded that none of the analyzed files complies with ISO and ANSI / ADA specifications, therefore quality control must be re-evaluated in order to obtain a standardization of the endodontic instruments that are available on the market.pt_PT
dc.identifier.tid202842789pt_PT
dc.identifier.urihttp://hdl.handle.net/10400.26/38870
dc.language.isoporpt_PT
dc.subjectLimas K-Filept_PT
dc.subjectNormas ISO e ANSI/ADApt_PT
dc.subjectDesign macroscópico e microscópicopt_PT
dc.subjectMEVpt_PT
dc.titleEstudo comparativo das características microscópicas e macroscópicas entre limas originais e limas de contrafaçãopt_PT
dc.typemaster thesis
dspace.entity.typePublication
rcaap.rightsopenAccesspt_PT
rcaap.typemasterThesispt_PT
thesis.degree.nameMestrado Integrado em Medicina Dentáriapt_PT