Publication
The establishment of computed tomography diagnostic reference levels in Portugal
| dc.contributor.author | Santos, Joana | |
| dc.contributor.author | Foley, Shane | |
| dc.contributor.author | Paulo, Graciano | |
| dc.contributor.author | McEntee, Mark | |
| dc.contributor.author | Rainford, Louise | |
| dc.date.accessioned | 2017-04-13T10:17:20Z | |
| dc.date.available | 2017-04-13T10:17:20Z | |
| dc.date.issued | 2014 | |
| dc.description.abstract | The aims of this study were to investigate the frequency of Portuguese computed tomography (CT) examinations, identify protocol application and establish diagnostic reference levels (DRLs). CT departments (n=5211) were surveyed nationally (June 2011-January 2012) and CT protocol information and dose data were collected, as were retrospective age-categorised paediatric CT data from three national paediatric centres. The proposed national CT DRLs (CTDIvol) for adults were 75, 18, 14, 18, 17, 36, 22, 27 and 16 mGy for head, neck, chest, abdomen, pelvis, cervical, dorsal, lumbar and joints, respectively. The levels for paediatric head and chest examinations were as follows: 48 and 2 mGy (newborns), 50 and 6 mGy (5 y olds), 70 and 6 mGy (10 y olds) and 72 and 7 mGy (15 y olds). A limited number of current paediatric protocols aligned to recommended international age categorisations. Portuguese DRLs were generally higher than European recommendations, suggesting potential for optimisation. The need for greater standardisation of age-categorised paediatric protocols was identified. | pt_PT |
| dc.description.version | info:eu-repo/semantics/publishedVersion | pt_PT |
| dc.identifier.doi | 10.1093/rpd/nct226 | pt_PT |
| dc.identifier.uri | http://hdl.handle.net/10400.26/18186 | |
| dc.language.iso | eng | pt_PT |
| dc.peerreviewed | yes | pt_PT |
| dc.publisher | Oxford University Press | pt_PT |
| dc.subject | Adolescent | pt_PT |
| dc.subject | Adult | pt_PT |
| dc.subject | Child | pt_PT |
| dc.subject | Preschool Child | pt_PT |
| dc.subject | Female | pt_PT |
| dc.subject | Humans | pt_PT |
| dc.subject | Infant | pt_PT |
| dc.subject | Newborn Infant | pt_PT |
| dc.subject | Male | pt_PT |
| dc.subject | Portugal | pt_PT |
| dc.subject | Radiation Dosage | pt_PT |
| dc.subject | Radiometry | pt_PT |
| dc.subject | Reference Standards | pt_PT |
| dc.subject | Risk | pt_PT |
| dc.subject | X-Ray Computed Tomography | pt_PT |
| dc.title | The establishment of computed tomography diagnostic reference levels in Portugal | pt_PT |
| dc.type | journal article | |
| dspace.entity.type | Publication | |
| oaire.citation.endPage | 317 | pt_PT |
| oaire.citation.issue | 3 | pt_PT |
| oaire.citation.startPage | 307 | pt_PT |
| oaire.citation.title | Radiation Protection Dosimetry | pt_PT |
| oaire.citation.volume | 158 | pt_PT |
| rcaap.rights | openAccess | pt_PT |
| rcaap.type | article | pt_PT |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- The establishment of computed tomography diagnostic reference levels in Portugal.pdf
- Size:
- 192 KB
- Format:
- Adobe Portable Document Format
License bundle
1 - 1 of 1
No Thumbnail Available
- Name:
- license.txt
- Size:
- 1.85 KB
- Format:
- Item-specific license agreed upon to submission
- Description: