Repository logo
 
Loading...
Thumbnail Image
Publication

Calibration, event reconstruction, data analysis, and limit calculation for the LUX dark matter experiment

Use this identifier to reference this record.
Name:Description:Size:Format: 
1.pdf7.7 MBAdobe PDF Download

Advisor(s)

Abstract(s)

The LUX experiment has performed searches for dark matter particles scattering elastically on xenon nuclei, leading to stringent upper limits on the nuclear scattering cross sections for dark matter. Here, for results derived from ${1.4}\times 10^{4}\;\mathrm{kg\,days}$ of target exposure in 2013, details of the calibration, event-reconstruction, modeling, and statistical tests that underlie the results are presented. Detector performance is characterized, including measured efficiencies, stability of response, position resolution, and discrimination between electron- and nuclear-recoil populations. Models are developed for the drift field, optical properties, background populations, the electron- and nuclear-recoil responses, and the absolute rate of low-energy background events. Innovations in the analysis include in situ measurement of the photomultipliers' response to xenon scintillation photons, verification of fiducial mass with a low-energy internal calibration source, and new empirical models for low-energy signal yield based on large-sample, in situ calibrations.

Description

Keywords

Citation

Research Projects

Organizational Units

Journal Issue